Towards in vivo biosensors for low-cost protein sensing A. Ramesh, F. Ren, P.R. Berger, P. Casal, A. Theiss, S. Gupta and S.C. Lee In vivo biosensing requires stable transistor operation in high-salt concentration bodily fluids while exhibiting impermeability to mobile alkali ions that would otherwise render the metal-oxide-semiconductor (MOS) threshold voltage to drift. Metal oxide semiconductor capacitor structures using Al2O3 as the gate dielectric were soaked in a sterile physiological buffer solution (PBS) up to 24 hours and for thicknesses from 100 to 10 nm. The triangular voltage sweep technique characterised alkali ion penetration, and measured no detectable alkali ions for the Al2O3 capacitors. By contrast, the dose of alkali ions in silicon dioxide MOS capacitors steadily increased with increasing soak times in the PBS solution.

metal. Finally, aluminium metal was deposited on the backside of the wafer to complete the capacitor fabrication, followed by a postmetallisation anneal at 450°C for 10 min in nitrogen ambient. The ALD Al2O3 dielectric constant was calculated to be 8.65 from C-V measurements with a thickness of 103 nm. gate

VGS RE

metal dielectric oxide

positive analyte receptor

p Si-substrate backside contact

metal D

S n+

functionalised surface

n+

dielectric oxide VDS

Fig. 1 Field effect transistor protein biosensor

Thermally grown silicon oxide (SiO2) was used as the control sample. Dry silicon oxide was grown in an atmospheric tube furnace at 1050°C with an oxygen ambient followed by a 20 minute nitrogen anneal at the same temperature. The SiO2 oxide thickness was measured to be 116 nm with a calculated dielectric constant of 3.8. The in vivo physiological environment can be simulated by conducting experiments in physiological buffer solutions (pH 7.4, 0.15 M Na + , K + ). Natural in vivo protein environments contain comparable concentrations of alkali ions at a similar pH. Hence, immunity of transistor electrical response to these environments serves as a viable proof of the applicability of Si-based FET sensors for in vivo measurements. Permeation of mobile charges into the oxide can be quantified using the triangular voltage sweep (TVS) method. The MOS sample is heated to a temperature ( ∼ 250°C) to increase ion mobility. The MOS capacitor is stressed for 5 minutes with a 1 MV/cm electric field across the oxide, moving all the mobile ions to one side of the capacitor plate. A triangular voltage ramp is subsequently applied to the gate of the capacitor. A quasi-static capacitance-voltage C-V measurement is performed. As the polarity of voltage changes, a displacement current due to mobile ion movement, results in a peak in the measured capacitance. Next, a high frequency C-V measurement is performed, where the ions do not have sufficient time to respond, and no significant peak due to mobile ions is observed. Using this as the baseline, the area between these two curves (high frequency and low frequency) is determined by integration to obtain the mobile ion charge density within the oxide. Finally, MOS capacitors with ALD Al2O3 and thermal SiO2 gate dielectrics were soaked in a PBS solution for varying amounts of time and subsequently measured by the TVS technique. 12 80

thermal 100nm SiO2

70 60

alkali ions, 1010 cm–2

Experiments: A representative field effect transistor (FET) protein sensor is illustrated in Fig. 1. For use as biosensors, the gate metal is replaced by a functionalised oxide surface with analyte-specific affinity reagents (receptors), leaving the gate ‘floating’ in direct contact with the solution being tested. Binding of charged analytes (protein to be detected) to these surface receptors results in a change in the charge induced in the channel, which manifests as a change in the drain current. Since a gate metal is absent, a voltage is applied to the electrolyte through a reference electrode (RE) to shift the baseline transistor bias condition and maximise transistor gain. A simple MOS capacitor can be effectively used to determine the presence of mobile ions, such as sodium (Na + ) ions, in the oxide. The typical structure of a MOS capacitor is as shown in the inset of Fig. 1. In this study, thermally grown silicon oxide (SiO2) is replaced by aluminium oxide (Al2O3) deposited by the atomic layer deposition (ALD) process. The Al2O3 deposition was carried out with trimethylaluminum (TMA) and water as precursors at 300°C using a Picosun SunaleTM reactor. The substrates used are moderately doped ( ∼ 1016 cm − 3) p-type silicon (100) wafers. The ALD pulsing sequence for one cycle was 0.1 s TMA pulse, 4 s N2 purge, 0.1 s H2O pulse, and 4 s N2 purge. Typical ALD deposition rates of the 0.8 Å/cycle were obtained. The samples were then annealed at 700°C in nitrogen ambient. Aluminium metal was deposited on the topside and patterned by photolithography and lift-off to obtain square electrodes with various areas of 275 × 275, 550 × 550, 1100 × 1100, 1650 × 1650 and 2200 × 2200 µm2. The square electrodes were alternatively designed, some with holes and slots to permit various levels of ion permeation. Control electrodes were also included with continuous metal and no

A sensing channel connects the source (S) and drain (D) with a reference electrode (RE). When a target protein binds to the receptor, it induces charges in the substrate, causing a change in the current flow between the source and drain Inset: Typical structure of a MOS capacitor used in this study

capacitance density, nF/cm2

Introduction: In the past two decades, research and development of biosensors and their applications has undergone remarkable growth [1]. Electrical biosensors have been applied to different target applications, including glucose, pH and protein, detection and measurement. Protein biosensors are of particular interest and importance in modern medicine for their role in the early detection and diagnosis of disease, for instance cancer. Protein biosensors based on different semiconductor materials have been explored, such as Si [2], AlGaN/GaN [3], and carbon nanotubes [4]. Si-based protein biosensors are low-cost and easily integratable onto a small chip atop a diagnostic needle complete with readout circuitry. However, Si-based protein biosensors suffer from long-term electrical drift and instability due to the diffusion of ions from high osmolarity biological buffers into the gate oxides. Long-term stable and low-cost Si-based in vivo protein biosensors are needed but their translation to in vivo sensing applications introduces big challenges. The typical in vivo physiological environment contains Na + and K + ions that can be incorporated into the dielectric oxide of the field effect transistor and contribute to mobile charge [5]. These mobile ions shift within the active device depending upon voltage, causing a variable drift in the transistor threshold voltage, resulting in inaccurate protein sensing. Hence, impermeability to mobile alkali ions is key for stable transistor operation. Alternate dielectric oxides with impermeability to alkali ions, or reduced ion mobility, can help mitigate this issue. In this Letter, alkali ion penetration is cited as a critical factor for threshold voltage instability in biosensors using SiO2 as a gate dielectric, and impermeability of ions through Al2O3 in high ion concentration (0.15M) physiological buffer solution is demonstrated. This allows the future realisation of low-cost Si-based in vivo biosensors.

0 min 30 min 60 min 90 min

50 40 30 20

–4

–2

0 2 voltage, V

a

4

thermal 100nm SiO2

10 8 6 4 2 30 40

50

60 70 time, min

80

90 100

b

Fig. 2 Triangular voltage sweep measurements of typical thermal SiO2 MOS capacitor at 250°C (Fig. 2a). Relationship between increases in alkali ions concentration with increasing soak times in PBS for control SiO2 based capacitor (Fig. 2b); line is joins of measured data

Results: Fig. 2a shows the result of TVS measurements for a typical 100nm SiO2 MOS capacitor. TVS measurements were conducted prior to dipping in PBS solution and after soaking in PBS for 30, 60, and 90 minutes. It should be noted that thermal SiO2 shows a mobile ion peak prior to soaking in PBS due to incorporation of some alkali ion contamination from the tube furnace during thermal oxidation. Additionally, as the soak time in the PBS solution is increased, a clear linear increase in the mobile ion peak is observed in SiO2. This indicates

ELECTRONICS LETTERS 28th March 2013 Vol. 49 No. 7

significant penetration of ions from the PBS solution into the SiO2 oxide. The area between consecutive curves quantifies the increased mobile charge (alkali ions) after each soak and is determined by numerical integration. Table 1 and Fig. 2b tabulate the increase in alkali ion penetration into SiO2 MOS capacitors with increasing soak times in the PBS solution.

Table 1: Relationship between increased alkali ion concentration into thermal SiO2 oxide ( ∼ 100 nm) and PBS soak times Time (min) 0 30 60 90 Δ[Alkali ions] × 1010 cm-2 0 1.77 3.69 10.87

capacitance density, nF/cm2

capacitance density, nF/cm2

The experiment was then repeated with a 100 nm-thick ALD Al2O3 gate dielectric. The results are depicted in Fig. 3a. No response due to alkali ion penetration is observed up to 24 hours soak time. This was consistent amongst all four electrode configurations, floating-gate, slots, holes, and opaque. Reduction in oxide thickness provides an additional benefit of increasing capacitance, hence increased sensitivity to analyte charge. Hence, MOS capacitors with reducing ALD Al2O3 oxide thicknesses were also fabricated and soaked in PBS as described above. Figs. 3b–d depict the TVS measurement results for 50, 25, 10 nm Al2O3 thicknesses, respectively. Again, no mobile ion response was observed for soak times in PBS up to 24 hours for any of these thinner Al2O3 oxide thicknesses, regardless of electrode perforations. 90

70 60 50 –2

100

100 nm ALD Al2O3

80

–1

120 100

0 voltage, V

1

2

80 –2

0 min 30 min 60 min 90 min 120 min 600 min 1440 min

–1

280 25 nm ALD Al2O3

240 200 160 120 –2

0 min 30 min 60 min 90 min 120 min 600 min 1440 min

–1

0 voltage, V

c

0 voltage, V

1

2

b

1

2

capacitance density, nF/cm2

capacitance density, nF/cm2

a 700

10 nm ALD Al2O3

600 500 400 300 200 100 –2

0 min 30 min 60 min 90 min 120 min 600 min 1440 min

–1

0 voltage, V

1

© The Institution of Engineering and Technology 2013 9 February 2013 doi: 10.1049/el.2012.4283 One or more of the Figures in this Letter are available in colour online. A. Ramesh, F. Ren and P.R. Berger (Department of Electrical and Computer Engineering, The Ohio State University, Columbus, OH 43210, USA) E-mail: [email protected] P. Casal, A. Theiss, S. Gupta and S.C. Lee (Department of Chemical and Biomolecular Engineering, The Ohio State University, Columbus, OH 43210, USA) References

50 nm ALD Al2O3

140

0 min 30 min 60 min 90 min 120 min 600 min 1440 min

Conclusion: Silicon in vivo protein biosensors suffer from long-term electrical drifting and instability due to the contamination of alkali ions from high osmolarity biological buffers. In this reported work, a Si based MOS capacitor with a high-k Al2O3 dielectric deposited by ALD has been fabricated. High-k dielectric layers not only prevent alkali ions diffusion from high osmolarity biological buffers into the gate oxides but also result in enhanced device sensitivity due to increased electrostatic coupling. MOS capacitors with Al2O3 gate dielectric show no measurable peak before and after soaking in the PBS solution, indicating no alkali ions penetration for various oxide thicknesses of 100, 50, 25, 10 nm. Thus, Si-based FETs with Al2O3 as their oxide dielectric are attractive candidates for the realisation of future low-cost biosensors for in vivo protein sensing.

2

d

1 Grieshaber, D., Mackenzie, R., Voros, J., and Reimhult, E.: ‘Electrochemical biosensors – sensor principles and architectures’, Sensors, 2008, 8, p. 1400 2 Wee, K.W., Kang, G.Y., Park, J., Kang, J.Y., Yoon, D.S., Park, J.H., and Kim, T.S.: ‘Novel electrical detection of label-free disease marker proteins using piezoresistive self-sensing micro-cantilevers’, Biosens. Bioelectron., 2005, pp. 1932–1938 3 Nicholson, T., Gupta, S., Wen, X., Wu, H., Anisha, R., Casal, P., Kwak, K.J., Bhushan, B., Berger, P.R., Lu, W., Brillson, L.J., and Lee, S.C.: ‘Rational enhancement of nanobiotechnological device functions illustrated by partial optimization of a protein-sensing field effect transistor’, J. Nanoeng. Nanosyst., 2010, 223, pp. 149–161 4 Wang, J.: ‘Carbon-nanotube based electrochemical biosensors: a review’, Electroanalysis, 2005, 17, (1), pp. 7–14 5 Kuhn, M., and Silversmith, D.: ‘Ionic contamination and transport of mobile ions in MOS structures’, J. Electrchem. Soc.,1971, 118, p. 966

Fig. 3 Triangular voltage sweep measurements of MOS capacitors with Al2O3 dielectric at 250°C for 100, 50, 25, and 10 nm oxide thicknesses (Figs. 3a–d, respectively) Note the absence of any time varying mobile ion peak superimposed on the baseline C-V waveform that instead shows a natural progression of depletion and inversion in the channel

ELECTRONICS LETTERS 28th March 2013 Vol. 49 No. 7

Towards in vivo biosensors for low-cost protein sensing

Feb 9, 2013 - centration bodily fluids while exhibiting impermeability to mobile ... Long-term stable and low-cost Si-based in vivo protein biosensors are.

240KB Sizes 1 Downloads 212 Views

Recommend Documents

CMOS MEMS Accelerometer for Long-Term In Vivo ...
Figure 1. In Vivo Wireless Bio-Sensing Network System. Figure 2. Implantable ... attached on the back of a laboratory mouse for the characterization. The mouse ...

The Need for Innovative In-Vivo Measurement ...
Apr 10, 2012 - present the requirements for whole brain emulation, and we introduce a feasible program ... system energy in terms of a Hamiltonian Hрx; ; u; tЮ ¼ ..... sources, receiving some through transmissions form external sources and some in

Environmental Biosensors
For environmental applications such systems depend primarily on the use .... relatively high development costs for single analyte systems, limited shelf and .... Bio Data. Sandeep. Sandeep K. Jha is presently working as a CSIR senior research ...

Biosensors International
Biosensors International. Bloomberg: BIG SP | Reuters: BIOS.SI. Refer to important disclosures at the end of this report. Muted growth. • 4Q14 and FY14 results missed expectations on weak royalty income, lower ASPs and weaker margins. • Expect mu

Effective Reranking for Extracting Protein-Protein ... - Semantic Scholar
School of Computer Engineering, Nanyang Technological University, ... of extracting important fields from the headers of computer science research papers. .... reranking, the top ranked parse is processed to extract protein-protein interactions.

Effective Reranking for Extracting Protein-Protein ... - Semantic Scholar
School of Computer Engineering, Nanyang Technological University, ... different models, log-linear regression (LLR), neural networks (NNs) and support vector .... reranking, the top ranked parse is processed to extract protein-protein ...

Biosensors International Alert
Aug 6, 2014 - -12.8%. Source: Deutsche Bank. Stock data. Market cap (SGDm). 1,430. Market cap (USDm). 1,145. Shares outstanding (m). 1,707.8. Major shareholders. NORGES BANK. (5.04%). Free float (%). 55. Avg daily value traded. (USDm). 5.6. Source: D

Extracting Protein-Protein Interactions from ... - Semantic Scholar
statistical methods for mining knowledge from texts and biomedical data mining. ..... the Internet with the keyword “protein-protein interaction”. Corpuses I and II ...

QoS Provision for Remote Sensing and Control in ...
mobile sensors. Sensory information, such as video and audio, will also be received and rendered by the control terminal. The control commands and sen- sory feedbacks, including possibly different types of media types, are referred to as supermedia [

Electrochemical Biosensors of nanostructured CuO ...
electrical and optical, photovoltaic devices [1-3]. To date, on the basis of the practical importance of CuO nanomaterials, well-defined CuO nanostructures with ...

Biosensors and Bioelectronics Hollow spherical ...
et al., 2003; Marinakos et al., 1999) as hard templates followed by the removal of the .... (0.1–0.5 mM; a–e, respectively) in PBS (pH 7) at a rotating Pt disk electrode modified ..... Supplementary data associated with this article can be found,