QA Maintenance and Inspection ViZion DR Detector Guide v2.0 1 Opal UAI v.2.3.4.x

QA Maintenance and Inspection Vizion DR Detector Guide For safety reasons, be sure to turn OFF the power of each instrument when the performed inspection indicates power to be off. Failing to power off instrument when requested may result in electric shock. When the instrument is cleaned, be sure to turn OFF the power of each instrument, and unplug the power supply cord from the AC outlet. Never use benzene, thinner or any other flammable cleaning agents, as this may result in fire or electric shock. The instrument must be repaired by a qualified engineer only. Failure to do so may result may result in fire, electric shock, or accident. I.

Environment of Use and Storage

Do not install the instrument in a location with the conditions listed below. Otherwise, it may result in failure or malfunction, fall or cause fire or injury.         

Close to facilities where water is used. Where it will be exposed to direct sunlight. Close to air-conditioner or ventilation equipment. Close to heat source such as a heater. Prone to vibration. Insecure place. Dusty environment. Saline or sulfurous environment. High temperature or humidity.

Do not place the storage case or instrument in a location with the conditions listed below.

II.



Where the cable of the sensor unit will be strongly pulled when the sensor unit is put into the case, otherwise, the cable may be damaged, resulting in fire or electric shock.



Where someone might get their foot caught in the cable of the sensor unit is put in the case. Otherwise they could trip over, resulting in injury

Power Supply 

The instrument’s cables are long, do not tangled during use.



Be careful not to get your feet caught in the cabling.

Customer Support VZDR-UG-QA Maintenance and Inspection ViZion DR Detector Guide-D-20141012-2.0

1.800.366.5343 – [email protected]

QA Maintenance and Inspection ViZion DR Detector Guide v2.0 2 Opal UAI v.2.3.4.x

III.

Handling 

Do not spill liquid or chemicals onto the instrument.



In cases where the patient is injured, do not allow the instrument to become wet with blood or other body fluids, as doing so may result in fire or electric shock. In such situation, protect the instrument with disposable covering as necessary.



Wipe the CFRP plate of the sensor unit with ethanol or glutaraldehyde solution to disinfect it each time a different patient uses the instrument, in order to prevent infection. For safety reasons turn off the power of each instrument when not in use.

IV.

V.

Maintenance and Inspection 

For safety reasons, be sure to inspect the instrument before using it. In addition, carry out a regular inspection at least once a year.



The instrument must be repaired by a qualified engineer only. If it is not repaired properly, it may cause fire, electric shock or accident.

Maintenance

VI.



Maintenance should be done by service provider



If the Detector Panel is defective, the detector will be returned as is to the manufacturer for repair



Clean the equipment with a dry soft cloth, or a soft cloth lightly moistened with mild detergent solution. Do not use any type of solvent, such as benzene



This equipment and accessories are to be disposed of safely after the life span of them and national regulation must be observed.



For safety reasons, be sure to inspect the instrument before using it. In addition, carry out a regular inspection at least once a year.



Arrange the detector and power supply link cable to prevent the damage of the cable’s rubber tube. For example, do not press the cable under the legs of the table or the people.

Inspection

In order to ensure that the instrument is used safely and normally, please be sure to inspect the instrument before use. If any problem is found during the inspection, please take measures indicated. If the problem still cannot be corrected, please contact ViZion DR Dealer representative or distributor. It is recommended that a record of the inspection be kept by making copies of the check lists in this section, or making a separate check list.

Customer Support VZDR-UG-QA Maintenance and Inspection ViZion DR Detector Guide-D-20141012-2.0

1.800.366.5343 – [email protected]

QA Maintenance and Inspection ViZion DR Detector Guide v2.0 3 Opal UAI v.2.3.4.x

For safety reasons, be sure to turn OFF the power of the detector when the following inspections are going to be performed. Otherwise, it may result in electric shock VII.

Daily Inspection  Before Turning ON the Power

 After Turning ON the Power 

VIII.

Perform the following inspection with the detector power supply.

Monthly Inspection Perform the following inspection periodically at least once a month. Contact your ViZion DR Dealer representative or distributor if there is any problem.

Phantom Inspection: 1. Check the performance of the instrument by performing exposures using a phantom or a resolution chart. Date: Results: Flat field Inspection: 1. Check the performance of the instrument by performing the flat field inspection instructions included in this document Date: Results: Customer Support VZDR-UG-QA Maintenance and Inspection ViZion DR Detector Guide-D-20141012-2.0

1.800.366.5343 – [email protected]

QA Maintenance and Inspection ViZion DR Detector Guide v2.0 4 Opal UAI v.2.3.4.x

IX.

Yearly Inspection Perform the following inspection periodically at least once a year. Contact your ViZion DR Dealer representative or distributor if there is any problem.

Phantom Inspection: 1. Check the performance of the instrument by performing exposures using a phantom or a resolution chart. Date: Results: Flat field Inspection: 1. Check the performance of the instrument by performing the flat field inspection instructions included in this document. Date: Results:

Flat Field Inspection of ViZion DR Instructions 1. Flat field inspection should be done on a monthly basis. 2. The flat field image should have pixel values within 10% of the median pixel value. 3. A deviation of more than 10% may denote the need for recalibration. Consult with a manufacture representative or service technician if pixel values deviate more than 10%. NOTE: If generator or X-ray tube power deviates from the original flat field kvP, SID or mAs previous pixel values will vary from previous flat field image values. 1. Disable ViZion DR enhancement a. Select Options from the acquisition screen. Figure 1 b. Select the Post-Processing tab from the configuration screen. Figure 2 c. Disable (uncheck)  Enable Auto Shutter

Figure 1

 Enable Enhancement  Enable Auto W/L

Customer Support VZDR-UG-QA Maintenance and Inspection ViZion DR Detector Guide-D-20141012-2.0

1.800.366.5343 – [email protected]

QA Maintenance and Inspection ViZion DR Detector Guide v2.0 5 Opal UAI v.2.3.4.x

d. Select the Developer Tab from the Configuration screen. Figure 3 e. Enable (check)  Show Pixel Value 2. Expose Flat Field Images a. 80 kVp, + or - 5 MAS @ 40SID b. 70 kVp, + or - 5 MAS @ 40SID c. 60 kVp, + or - 5 MAS @ 40SID

3. Inspect Pixel Values of Flat field image. Move the mouse pointer over the images to

Figure 2

display different pixel values

a. Mean ___________________ b. Median __________________ c. Average _________________ 4. Inspect Flat Field Image for artifacts (should have a homogenous grey scale look)

Note: The width for W/L setting for flat field images should be set to a value of at least 300 during visual inspection. Figure 3

Customer Support VZDR-UG-QA Maintenance and Inspection ViZion DR Detector Guide-D-20141012-2.0

1.800.366.5343 – [email protected]

Opal UAI v.2.3.4.x - 20/20 Imaging

The instrument must be repaired by a qualified engineer only. Failure to do so may result may result in fire, electric shock, or accident. I. Environment of Use and ...

792KB Sizes 5 Downloads 278 Views

Recommend Documents

OPAL - PDF COMBINED.pdf
All prices include VAT. All products are subject to availability. Some dishes may contain nuts or traces of nuts. Food allergies &. intolerances; before you order your food and drinks please speak to our staff if you have a food allergy or intoleranc

Opal Bar Menu.pdf
... cards accepted. There may be an optional service charge added to your bill. Page 1. Whoops! There was a problem loading this page. Opal Bar Menu.pdf.

Opal Bar Menu.pdf
Whoops! There was a problem loading more pages. Retrying... Opal Bar Menu.pdf. Opal Bar Menu.pdf. Open. Extract. Open with. Sign In. Details. Comments.

Opal food menu .pdf
Page 1 of 60. Bohol Profile. Bohol. Basic Facts. Geographic Location Bohol is nestled securely at the heart of the Central. Visayas Region, between southeast of Cebu and southwest. of Leyte. Located centrally in the Philippine Archipelago, specifical

3 Opal - Jennifer L. Armentrout.pdf
Page 1. Page 2 of 1,525. Whoops! There was a problem loading this page. 3 Opal - Jennifer L. Armentrout.pdf. 3 Opal - Jennifer L. Armentrout.pdf. Open. Extract.

DIGITALEUROPE Vision 2020
Chapter 2: The transformational power of digital technologies: sector and case examples. 50. Chapter 3: ...... operation of major new renewable energy sources.

2020.pdf
There was a problem previewing this document. Retrying... Download. Connect more apps... Try one of the apps below to open or edit this item. 2020.pdf. 2020.

Opal Card Reissue 2017.pdf
Whoops! There was a problem loading more pages. Opal Card Reissue 2017.pdf. Opal Card Reissue 2017.pdf. Open. Extract. Open with. Sign In. Main menu.

3. Opal-Jennifer L. Armentrout.pdf
b) Tính Cos DAB . c) Kẻ OM ⊥ BC ( M ∈ AD) . Chứng minh 1. BD DM. DM AM. − = d) Tính diện tích phần hình tứ giác OBDM ở bên ngoài nửa đường tròn (O).

Descargar libros opal carew
Descargar libros opalcarew.descargar la biblia gratis paracelularen pdf. ... android gratis.descargar tango roxanne.descargar fuentes gratiszurich.descargar ...

DIGITALEUROPE Vision 2020
Chapter 1: The transformational power of digital technologies. 34. Chapter 2: The ..... network-based interactive digital services – sometimes referred to as 'The ...

3. Opal-Jennifer L. Armentrout.pdf
Page 3 of 380. 3. Índice. Sinopsis. Capítulo 1. Capítulo 2. Capítulo 3. Capítulo 4. Capítulo 5. Capítulo 6. Capítulo 7. Capítulo 8. Capítulo 9. Capítulo 10. Capítulo ...

1_slcc_draft_strategy_2016-2020.pdf
organisation at peak efficiency and effectiveness, and one demonstrating the early impact of ambitious work to. deliver even greater value to consumers and to lawyers. To get there, we are seeking your views on a refreshed approach to our work. We be

2020 RESULTS 20M.pdf
2020 RESULTS - Page 1 of 5. POS BIB Firstname Surname CLUB TIME. 25 209 Jacob Hayes. M Bitton Road Runners 1:59:09. 1. M V40. 30 255 Ryan Mcflynn. M Mmrt 2:01:26. 1. M SEN. 31 314 Stefan Simms. M Mmrt 2:01:43. 2. M V40. 36 287 Toby Pearson. M Thames

India Vision 2020
Dec 5, 2002 - This initiative brought together over 30 experts from different fields. ..... The overall growth in demand for all forms of fuel will mirror the growth in the ... Greater reliance on renewable energy sources offers enormous economic, ..

Frayser 2020.pdf
Good data and information will help drive our decision-making. Page 3 of 108. Frayser 2020.pdf. Frayser 2020.pdf. Open. Extract. Open with. Sign In. Main menu.

e 2020 electives.pdf
Sign in. Loading… Page 1. Whoops! There was a problem loading more pages. e 2020 electives.pdf. e 2020 electives.pdf. Open. Extract. Open with. Sign In.

2016 School Opal Card Fact Sheet.pdf
2016 School Opal Card Fact Sheet.pdf. 2016 School Opal Card Fact Sheet.pdf. Open. Extract. Open with. Sign In. Main menu.

ompi onet opal reg stci stci−rte.h - GitHub
#define orte_rml stci_rml with stci declaring same functionality. Register BTL functions in the STCI. Callbacks are used in STCI and may be ignored in ORTE.

DBS Comprehensive Plan - 2020.PDF
Regulate the quality of new development through appropriate site plan review criteria;. b Provide for recreational lands and facilities to meet the Level of Service ...

RENSTRA FKIP 2015-2020.pdf
dipandang perlu adanya suatu rencana strategis yang. merupakan landasan pada penyusunan rencana operasional. pencapaian visi, misi, tujuan pendidikan ...